In the first two articles, we explored the trend of intelligent weighing sensors and the physical survival rules under harsh working conditions. But in real industrial sites, it is often the invisible electromagnetic interference that breaks down a weighing system, not because the steel is not hard enough or the shell is not thick enough.
The instrument numbers keep jumping and cannot be read at all. "" As soon as the frequency converter next to it starts, the scale data will skyrocket instantly. "" The weight captured by the PLC always has a deviation of a few tenths of a kilogram. More than 90% of these problems are caused by electrical noise, grounding circuits, or improper parameter settings. This article will deeply analyze the signal interference mechanism and practical troubleshooting of weighing systems from a micro electrical perspective.
一, The fragility of signals: why micro volt level interference can lead to big mistakes?
To understand interference, one must first understand how fragile the signal from the weighing sensor is.
A standard analog weighing sensor typically has a sensitivity of 2.0 mV/V. If a 10V excitation voltage is provided to it, then at full load (e.g. 30 tons), its full-scale output signal is only 20mV.
If the range of the sensor is 30 tons (30000 kilograms), it means that a weight change of 1 kilogram corresponds only to a voltage change of 0.67 microvolts (μ V).
Any small electromagnetic field changes, induced electromotive force on the circuit, or even static electricity generated by metal friction can potentially add noise to this weak 20mV signal. When noise overwhelms real weight changes, accuracy becomes impossible to talk about. That's why industrial weighing heavily relies on shielding, filtering, and proper grounding.
二, The Three Major Sources of Interference and Engineering Countermeasures
1. Power equipment interference (frequency converter and welding machine)
Phenomenon: When the workshop crane or conveyor motor starts, the weighing data instantly overflows or jumps violently.
Truth: When the IGBT high-frequency switch works, the frequency converter will generate strong electromagnetic radiation and high-frequency harmonics. If the sensor cable runs parallel to the power line, these radiations will be directly coupled into the signal line.
Countermeasure:
Physical isolation: The signal line of the sensor must be routed separately from the power line, and parallel laying is strictly prohibited. If it is necessary to cross, be sure to cross vertically to minimize the sensing area.
Shielding layer single ended grounding: This is the most easily misconnected wire on site. Sensor cables all have aluminum foil or woven mesh shielding layers, which must and can only be grounded at a single point on the instrument end. If both ends are connected, once there is a potential difference between the two ends, a "grounding loop" will be formed, generating greater common mode interference than not grounding.
2. Welding current: the "instantaneous killer" of sensors
Phenomenon: After the factory maintenance was completed, a certain sensor completely failed, and the input impedance was measured with a multimeter to be open circuit (infinite).
Truth: When performing welding operations, if the grounding wire of the welding machine is not placed on the scale body near the welding point, hundreds of amperes of welding current will find the shortest path and directly pass through the strain gauges and internal leads of the sensor, instantly burning it out.
Countermeasure: Strictly establish operating procedures. Before welding, it is necessary to disconnect the sensor from the instrument; If it is not turned on, the ground wire of the welding machine must be directly clamped onto the metal part that needs to be welded to ensure that the sensor is not in the current circuit.
3. Common mode voltage and leakage current
Phenomenon: The sensor is not under force, but the instrument displays a huge zero drift.
Truth: In large storage tanks or long-distance weighing scales, the leakage current of power equipment will flow into the ground along the metal structure. If the ground potential of the instrument power supply is not consistent with the ground potential of the scale body where the sensor is located (there is a potential difference), a common mode voltage will be generated in the signal circuit.
Countermeasure: Use high-quality digital weighing instruments (usually with better common mode rejection ratio CMRR); In strong interference environments, switch to digital weighing sensors or install signal isolators to cut off the conduction path of potential difference through photoelectric isolation.
三, On site troubleshooting three board axe: practical guide for multimeter operation
When suspecting a sensor malfunction, do not replace it blindly. Use a multimeter to perform the following three steps of troubleshooting (power off operation is required):
Measure input/output impedance: The input impedance (usually 400 Ω or 750 Ω) and output impedance (usually 350 Ω or 700 Ω) of a normal sensor should be around the standard value. If the impedance abnormally increases (open circuit) or decreases (short circuit/water ingress), the sensor is damaged.
Measure insulation resistance: Use a megohmmeter (shaker) to measure the resistance between the sensor bridge and the housing (shielded wire). Normally, it should be greater than 5000 M Ω (50VDC). If the insulation resistance is extremely low, it indicates that water and moisture have entered the interior, causing leakage in the bridge circuit.
Measurement of unbalanced voltage (weighing signal): Under no-load conditions, measure the output terminal of the sensor in millivolt range (excitation between EX+/EX -, signal between SIG+/SIG -). Under the same excitation voltage, the unbalanced voltage of each sensor in the same system should be basically the same. If the deviation is too large, it indicates that the sensor is damaged or plastic deformation occurs in the strain zone.
四, Solution from the source: collaboration between digital filtering algorithms and full inspection quality control
In addition to external hardware shielding, the filtering ability of software and firmware is also crucial in combating electrical interference.
Modern weighing instruments and industrial systems (such as PLCs) will integrate digital filtering algorithms. From simple sliding average filtering to advanced adaptive filtering (such as the application of Kalman filtering in dynamic weighing), the essence is to find a balance between "filtering out high-frequency noise" and "maintaining fast response to weight changes". For dynamic high-speed sorting (Checkweigher), excessive filtering delay can lead to missed detections; For static material tanks, stronger filtering results in more stable data.
However, no matter how powerful the filtering algorithm of the instrument is, the premise is that the dispersion of the sensor itself should be small and the internal consistency should be good. This is also the engineering significance of Shandong Leheng and other source factories adhering to "100% full inspection and shipment" and "full process precision calibration".
100% full inspection not only detects the final weight output, but also includes rigorous screening of the bridge impedance and insulation performance of each sensor. Only by eliminating defective products with internal micro cracks or insulation defects from the production end can sufficient margin be left for the system's anti-interference ability in harsh environments filled with frequency converters, static electricity, and long cables.
System level thinking is the ultimate solution to solving problems
An industrial weighing system is a complex entity composed of physical structures (sensors, scales), electrical links (cables, junction boxes), and digital processing (instruments, algorithms).
When there is a data jump on site, don't just focus on the "possibly broken" sensor. Go check the position of the frequency converter, check the grounding of the shielded wire, and measure if there is any leakage current in the scale body. Only by understanding the underlying logic from microvolt level signals to electromagnetic coupling can you move from "passive maintenance" to "active design" and ensure the rock solid strength of the weighing system.